Trial Number 3

Trial Purpose:

To eliminate the use of N-Propyl Bromide in cleaning operations

Date Run:

04/11/2016

Experiment Procedure:

Stainless steel coupons were weighed to determine the initial weights. The bottom third of the coupons were soiled with Sonnen B-200L oil using a swab. Three coupons were immersed at room temperature in the Fluosolv CX, the other remaining coupons were immersed at room temperature DuPont Vertrel SION. Each set of coupons were cleaned for five minutes each. Once the coupons were removed from the cleaner, the coupons were hung to air dry fifteen minutes. The coupons were reweighed, and percentage calculations were completed.

Trial Results:

The Fluosolv CX was inefficient at removing the Sonnen B-200L oil from the stainless-steel coupons, with an average percentage removal of 61.48%. However, the DuPont Vertrel SION was very efficient at removing the Sonnen B-200L oil, its average removal rate was 98.86%.

Cleaner Initial wt of cont. Final wt of cont. %Cont Removed
Fluosolv CX 0.0100 0.0039 61.00
  0.0142 0.0047 66.90
  0.0122 0.0053 56.55
Vertrel Sion 0.0171 0.0000 97.08
  0.0198 0.0001 99.49
  0.0081 0.0000 100.00

Success Rating:

Results successful using TACT (time, agitation, concentration, and temperature, as well as rinsing and drying) and/or other cleaning chemistries examined.

Conclusion:

The Fluosolv CX was inefficient at removing the Sonnen B-200L oil from the stainless-steel coupons, with an average percentage removal of 61.48%. However, the DuPont Vertrel SION was very efficient at removing the Sonnen B-200L oil, its average removal rate was 107.08%.