Trial Purpose:

Request for approval & information

Date Run:

09/12/1994

Experiment Procedure:

We are shipping out the samples for our cleaning comparison to be performed by the lab. We have made a couple of changes in the items:
1. The nickel piece is a large flat disc.
2. The monel piece is of two different configurations.
The samples will be labeled as follows:
Copper - #’s 1, 2, 3, 4 &5
Kovar - #’s 1, 2, 3, 4 & 5
Ceramic - #’s 1, 2, 3 & 4
Nickel - #’s 1, 2, 3 &4
Tungsten - #’s 1, 2, 3 & 4
Molybdenum - #’s 1, 2, 3 &4
Monel - #’s 1, 2 &3
Quartz - #’s 1, 2 &3

BACKGROUND: There are currently four chemical labs in our facility that process similar or identical materials independent of the others due to the fact that they belong to separate business elements. In most cases the cleaning/processing is slightly different for each department. We are currently in an effort to reduce our wastes produced as a facility. By doing so we are trying to consolidate our chemical lab processing.
PLAN: Send you (SCL) samples of selected parts (based upon the material) that have been cleaned by each of our departments utilizing that material. Then accompanying each part/material selection will be a sample that has been cleaned by a chosen new method. The new methods are ones that do not use chlorinated solvents, ozone depleting chemicals, are non-hazardous, etc.
We wish to compare the current methods to the new methods not only to prove in the new methods but to establish a baseline to compare them to.
Could you please look this information over and get back to me with your concerns and possibly a time frame that this work can be completed. If you need further information please do not hesitate to call. We will be speaking to you in the near future.

Trial Results:

OSEE readings were taken for all samples except for the Molybdenum due to its small size. For each sample, OSEE readings were taken at 10 second intervals for one minute. Due to the nature of Quartz readings, readings were not able to be taken every 10 seconds. The peak readings were taken instead. The average reading for the Quartz was 159.57 while the average reading for the Quartz #1 sample was 38.50. The results from all other OSEE readings are shown below.

OSEE TESTS

     OSEE TESTS GONIOMETER
SAMPLE TIME (S) OSEE1 OSEE2 OSEE3 OSEE4 AVG OSEE ANGLE
COPPER 3 0 70 56 42   56.00 70.00
  10 56 50 35   47.00 60.00
SENSOR GAIN  20 51 46 32   43.00 75.00
SET AT 7 30 47 44 30   40.33 80.00
  40 45 42 28   38.33 80.00
  50 42 40 27   36.33  
  60 41 39 26   35.33  
            AVERAGE 73.00
               
               
    OSEE TESTS  GONIOMETER
SAMPLE TIME (S) OSEE1 OSEE2 OSEE3 OSEE4 AVG OSEE ANGLE
COPPER #5 0 378 288 316   327.33 83.00
  10 310 258 279   282.33 85.00
SENSOR GAIN  20 288 243 259   263.33 80.00
SET AT 7 30 278 235 247   253.33 85.00
  40 272 228 239   246.33  
  50 262 225 232   239.67  
  60 255 219 225   233.00  
            AVERAGE 83.25
               
               
               
    OSEE TESTS GONIOMETER
SAMPLE TIME (S) OSEE1 OSEE2 OSEE3 OSEE4 AVG OSEE ANGLE
KOVAR #3 0 102 85 70   85.67 80.00
  10 93 82 70   81.67 70.00
SENSOR GAIN  20 81 77 66   74.67 70.00
SET AT 7 30 73 76 67   72.00 68.00
  40 70 75 65   70.00  
  50 69 72 65   68.67  
  60 66 72 63   67.00  
            AVERAGE 72.00
               
               
    OSEE TESTS GONIOMETER
SAMPLE TIME (S) OSEE1 OSEE2 OSEE3 OSEE4 AVG OSEE ANGLE
KOVAR #5 0 53 53 46   50.67 30.00
  10 44 52 46   47.33 35.00
SENSOR GAIN  20 41 52 45   46.00 20.00
SET AT 7 30 39 50 44   44.33 30.00
  40 37 49 44   43.33  
  50 36 48 44   42.67  
  60 35 47 43   41.67  
            AVERAGE 28.75
               
               
    OSEE TESTS GONIOMETER
SAMPLE TIME (S) OSEE1 OSEE2 OSEE3 OSEE4 AVG OSEE ANGLE
CERAMIC #3 0 31 24 19   24.67  
  10 17 10 17   14.67  
SENSOR GAIN 20 15 8 12   11.67  
SET AT 0 30 14 7 10   10.33  
  40 13 7 9   9.67  
  50 13 7 9   9.67  
  60 13 7 9   9.67  
            AVERAGE #DIV/0!
               
               
    OSEE TESTS GONIOMETER
SAMPLE TIME (S) OSEE1 OSEE2 OSEE3 OSEE4 AVG OSEE ANGLE
CERAMIC #4 0 25 25 15   21.67  
  10 21 11 7   13.00  
SENSOR GAIN 20 17 7 5   9.67  
SET AT 0 30 16 7 5   9.33  
  40 16 7 5   9.33  
  50 15 7 5   9.00  
  60 15 6 5   8.67  
            AVERAGE #DIV/0!
               
               
               
    OSEE TESTS  GONIOMETER
SAMPLE TIME (S) OSEE1 OSEE2 OSEE3 OSEE4 AVG OSEE ANGLE
NICKEL #3 0 182 159 205   182.00 70.00
  10 245 233 210   229.33 65.00
SENSOR GAIN  20 193 196 177   188.67 70.00
SET AT 9 30 180 178 160   172.67  
  40 169 165 152   162.00  
  50 165 157 142   154.67  
  60 157 151 137   148.33  
            AVERAGE 68.33
               
               
    OSEE TESTS  GONIOMETER
SAMPLE TIME (S) OSEE1 OSEE2 OSEE3 OSEE4 AVG OSEE ANGLE
NICKEL #1 0 108 125 245   159.33 85.00
  10 759 681 876   772.00 82.00
SENSOR GAIN  20 694 630 802   708.67 85.00
SET AT 9 30 646 610 767   674.33 80.00
  40 612 583 748   647.67  
  50 591 577 735   634.33  
  60 579 555 717   617.00  
            AVERAGE 83.00
               
               
    OSEE TESTS GONIOMETER
SAMPLE TIME (S) OSEE1 OSEE2 OSEE3 OSEE4 AVG OSEE ANGLE
TUNGSTON #3 0 126 184 195 108 153.25 40.00
  10 270 462 262 260 313.50 38.00
SENSOR GAIN  20 251 360 208 229 262.00 45.00
SET AT 8 30 244 333 195 215 246.75 39.00
  40 235 312 185 207 234.75 40.00
  50 229 300 176 197 225.50 42.00
  60 225 288 169 191 218.25 45.00
            AVERAGE 41.29
               
               
    OSEE TESTS GONIOMETER
SAMPLE TIME (S) OSEE1 OSEE2 OSEE3 OSEE4 AVG OSEE ANGLE
TUNGSTON #4 0 262 122 184 181 187.25 17.00
  10 502 210 322 253 321.75 15.00
SENSOR GAIN 20 455 180 279 224 284.50 27.00
SET AT 8 30 418 174 263 214 267.25 23.00
  40 399 168 249 206 255.50 22.00
  50 384 164 238 198 246.00 20.00
  60 367 160 230 193 237.50 20.00
            AVERAGE 20.57
               
               
               
    OSEE TESTS  GONIOMETER
SAMPLE TIME (S) OSEE1 OSEE2 OSEE3 OSEE4 AVG OSEE ANGLE
MONEL #2 0 47 132     89.50 83.00
  10 191 187     189.00 80.00
SENSOR GAIN 20 155 169     162.00 85.00
SET AT 8 30 141 160     150.50 80.00
  40 135 150     142.50  
  50 126 140     133.00  
  60 120 134     127.00  
            AVERAGE 82.00
               
               
    OSEE TESTS GONIOMETER
SAMPLE TIME (S) OSEE1 OSEE2 OSEE3 OSEE4 AVG OSEE ANGLE
MONEL #3 0 64 91     77.50 70.00
  10 89 167     128.00 62.00
SENSOR GAIN 20 80 150     115.00 72.00
SET AT 8 30 76 138     107.00 75.00
  40 73 131     102.00  
  50 71 126     98.50  
  60 69 122     95.50  
            AVERAGE 69.75

Success Rating:

A cleanliness study, addressing only various analytical techniques.

Conclusion:

Goniometry tests were done on every sample except for the ceramic, quartz and molybdenum because of the shape of the samples.