Trial Number 4

Trial Purpose:

To evaluate ozone levels over time for CleanCore Aqueous Ozone solution (ozonated water) Times = 0, 2 & 4 hrs and compare different methods of measuring aqueous ozone.

Date Run:

11/18/2016

Experiment Procedure:

The CleanCore Aqueous Ozone kiosk machine was turned on and run to generate ozonated water. Ozonated water was collected into the supplied CleanCore spray bottle.

Prior to running any cleaning trials, the system was used to compare the analytical devices to determine how comparable values would be from each system. The comparison of ORP (mV) readings between Clean Core Aqueous Ozone On-Demand Ati Q45H In-Line Meter and HANNAH HI 98121 Hand-held Meter was done by measuring ORP values after the system was operating at a steady state. Values were recorded and compared.

Properties of tap water and the ozonated water (directly from the spray hose and the CleanCore Spray Bottle) were measured and recorded throughout the procedure, as necessary, including temperature, ORP values (mV), dissolved ozone levels (ppm), pH and water hardness.

Instrumentation used for measurements:
on-machine: dissolved ozone meter - ATi Q45H (ozone in ppm and temp in °F); ORP meter - Black Stone BL982411 ORP Controller (ORP in mV)
handheld instruments: Hanna HI 98121 meter (ORP & temp in °C); Chemetrics Meter with vacu-vials (dissolved ozone in ppm)

At the appropriate time interval, based on the age of the ozonated solution in the CleanCore Spray Bottle (t= 0 hr, t=2 hr, t=4 hr), the cleaning solution from the CleanCore Spray Bottle was analyzed for ozone level and ORP values.

Water hardness was measured to be around 150 ppm in all trials.

Chemistries Evaluated: CleanCore Aqueous Ozone Solution (at t=0 hr, t=2 hr, t=4 hr) generated by CleanCore Aqueous Ozone Kiosk model machine, stored in CleanCore Spray Bottle;

Trial Results:

The initial data for the two ORP units were tabulated and a correlation equation was established. The resulting equation was found to have an R^2 value of 0.9978 indicating a very good fit of the equation to the data collected. 

Table of ORP values in mV   Date: Oct 5 2016

In-Line Hand Held    
ATi Q45H HANNAH-HI9812  Temp °F Temp °C
283 243 59.3 15.2
278 242 59.3 15.2
265 233 58.6 14.8
266 234 58.6 14.8
267 233 58.6 14.8
290 233 59.1 15.1
306 237 59.2 15.1
345 245 59.3 15.2
389 265 59.4 15.2
419 282 59.4 15.2
453 302 59.5 15.3
483 316 59.5 15.3
509 337 59.6 15.3
530 361 59.7 15.4
553 380 59.8 15.5
582 404 60.0 15.5
602 422 60.2 15.7
617 440 60.3 15.7
635 470 60.7 15.9

In each of the four cleaning trials (bathroom -B, DCC 17 -D, glass -G and Hucker’s soil –H), the ORP readings taken from the CleanCore spray bottle dropped down to an average of 357 mV (tap water average 242 mV), after two hours the average were nearly identical to tap water (264 to 231 mV) and at four hours the difference was even smaller (238-225 mV)

Date: 9 Nov 2016          
Reading from CleanCore HANNA Handheld        
In-Line Instruments Meter Temp wall meter Temp hand held    
Dissolved Oxygen          
(ppm) ORP (mV) ORP (mV) Calculated Deg F (C) Deg F (C)
1.164 917 908 860 62.8 (17.1) 61.9 (16.6)
1.111 919 897 864 62.5 (16.9) 61.9 (16.6)
1.111 919 912 864 62.4 (16.9) 61.7 (16.5)
1.177 919 908 864 62.3 (16.8) 61.7 (16.5)
Date: 10 Nov 2016          
Reading from CleanCore HANNA Handheld        
In-Line Instruments Meter Temp wall meter Temp hand held    
Dissolved Oxygen          
(ppm) ORP (mV) ORP (mV) Calculated Deg F (C) Deg F (C)
1.063 941 909 908 69.4 (20.8) 70.0 (21.1)
1.214 943 907 912 69.4 (20.8) 70.0 (21.1)
0.992 942 910 910 69.4 (20.8) 70.2 (21.2)
1.051 942 900 910 69.4 (20.8) 70.2 (21.2)
Date: 15 Nov 2016          
Reading from CleanCore HANNA Handheld        
In-Line Instruments Meter Temp wall meter Temp hand held    
Dissolved Oxygen          
(ppm) ORP (mV) ORP (mV) Calculated Deg F (C) Deg F (C)
0.996 933 880 892 67.1 (19.5) 66.4 (19.1)
1.003 934 871 894 66.9 (19.4) 66.2 (19.0)
1.006 934 920 894 66.8 (19.3) 66.2 (19.0)
1.007 935 883 896 66.7 (19.3) 66.0 (18.9)

For the dissolved ozone in parts per million, the initial value from the unit ranged from 1.01 – 1.24 ppm. When measured from the CleanCore spray bottle at time zero, the ppm levels were 0.19-0.46 ppm for the four trials. After two hours the ppm levels were ranged from 0.03-0.1, at or equal to tap water levels. It appears that the use of the supplied spray bottle reduces the ozone level in solution under these laboratory conditions.

ORP - mV t=0 hrs t=2 hrs t=4 hrs Deg F Deg C
Hannah meter          
Bathroom (spray) 254 224 223 71.8 22.1
Hannah meter DCC 17          
(spray) 365 342 233 72.0 22.2
Hannah meter Glass          
(spray) 402 248 274 70.8 21.6
Hannah meter Huckers          
(spray) 409 244 223 75.1 23.9
Hannah meter Tap water          
Bathroom run 205 211 205 72.5 22.5
Hannah meter Tap water          
DCC 17 run 262 238 242 71.7 22.1
Hannah meter Tap water          
Glass run 214 242 242 69.8 21.0
Hannah meter Tap water
Hucker run 287 236 212 73.5 23.1
Unit meter Bath 930     73.9 23.3
Unit meter DCC 930     70.2 21.2
Unit meter Glass 930     68.1 20.1
Unit meter Huckers 930     74.3 23.5

A similar trend was shown for the Vacu Vial (VV) measurements.

Dissolved Ozone - PPM t=0 hrs t=2 hrs t=4 hrs Deg F Deg C
VV - Bathroom (spray) 0.21 0.05 0.01 71.8 22.1
VV - DCC 17 (spray) 0.19 0.03 0.02 72.0 22.2
VV - Glass (spray) 0.23 0.1 0 70.8 21.6
VV - Huckers (spray) 0.46 0.04 0.01 75.1 23.9
VV - Tap water B 0.04     72.5 22.5
VV - Tap water D 0.04     71.7 22.1
VV - Tap water G 0.02     69.8 21.0
VV - Tap water H 0.03     73.5 23.1
meter B 1.044 - 1.23     73.9 23.3
meter D 1.103 - 1.149     70.2 21.2
meter G 1.141 - 1.236     68.1 20.1
meter H 1.010 - 1.200     74.3 23.5
VV - Fill hose B 0.85     71.8 22.1
VV - Fill hose D 0.76     72.0 22.2
VV - Fill hose G 0.89     70.8 21.6
VV - Fill hose H 1.06     75.1 23.9

Success Rating:

A cleanliness study, addressing only various analytical techniques.

Conclusion:

Each of the analytical tools used to measure ozone levels in the supplied unit showed that the level of ozone in solution drastically dropped when sprayed through the supplied CleanCore bottle and over the four hour testing period.

When comparing the ORP meters from the wall unit and hand held device showed that even though both systems had different mV values, the hand held unit could be used in the field as an assessment tool supplying relative ORP values as compared to the wall system. For the Vacu-Vial comparison, values were consistently lower than the wall system values as well. However, the vials could still be used to provide in-field measurement in the same way as the hand held ORP unit.

Both surrogate measurement systems demonstrated the downward trend of ozone levels overtime. Both showed that the use of the spray unit degraded ozone levels significantly.